Specifications


Item

Specifications

Applicable plates

MTP format MALDI plates (96 well, 384 well): 9 plates at max
Applied Biosystems (AB) format MALDI plates (192 well): 18 plates at max

Applicable flow rate for fractionation

0.1 μL/min to 50 μL/min (Total flow after mixing matrix)

Spotting interval

5 to 60 seconds in increments of 1 second

Matrix flow rate

0.1 μL/min to 50 μL/min in increments of 0.1 μL/min

Spotting modes

Continuous mode (spotting of certain quantity)
Time specific mode (continuous spotting only on predetermined sections)

Drain

Discharged to a draining vial attached to the plate holder

Cleaning of the probe

A cleaning process using a dedicated flow channel (Matrix deposited to the tip of the probe can be cleaned.)

Dimensions and weight

W420 x H550 x D500 mm
40 kg

Ambient temperature and humidity ranges

Temperature: 16 to 26 °C
Humidity: 40 to 70% (no condensation)

Plot direction

Compatible with AXIMA, and Applied Biosystems products
Straight or zigzag

Camera images

max 720 x 480, rate 30 fps (when the capture board 228-46043-91 is used)