Print this page

 
Shimadzu - Solutions for Science since 1875 SHIMADZU EUROPA GmbH
Global Links
You are here: >> Skip Navigation LinksHome > Products > Spectroscopy > FTIR > IR Prestige-21 > Specifications

Spectroscopy

IR Prestige-21
Specifications


Interferometer Michelson interferometer (30 degree incident angle)
Advanced Dynamic alignment system
Sealed and desiccated interferometer with an automatic dryer
Beam splitter Germanium-coated KBr plate for Middle IR (Standard)
Germanium-coated CsI plate for Middle/Far IR (Optional)
Silicon-coated CaF2 plate for Near IR (Optional)
Light source Air Cooled Ceramic for Middle/Far IR with 3 years guarantee (Standard)
Tungsten lamp for Near IR (Optional)
Detector DLATGS detector with Temperature Control for Middle/Far IR (Standard)
MCT (Hg-Cd-Te) detector with Liquid Nitrogen Cooling for Middle IR (Optional)
InGaAs detector for Near IR (Optional)
Wavenumber range 7,800 - 350 cm-1
12,500 - 240 cm-1 (Optional)
Resolution 0.5, 1, 2, 4, 8, 16 cm-1 (Middle/Far IR)
2, 4, 8, 16 cm-1 (Near IR)
S/N ratio 40,000: 1 and higher
(Peak-to-peak, 4 cm-1 resolution, approx. 2100 cm-1, 1 minute scan)
Operation System Microsoft Windows 2000/XP
Interface between PC and FTIR IEEE 1394
Hardware Monitoring Self Diagnostics, Status Monitor
Validation Program complies with Japanese Pharmacopoeia/European Pharmacopoeia/ASTM
Data processing Addition, Multiplication, Abs to %T conversion, Normalization, Baseline correction, Log conversion, Smoothing, Derivative, ATR correction, Kubelka-Munk conversion, Kramers-Kronig analysis, Wavenumber/wavelength conversion, Peak detection, Peak area calculation, Film thickness calculation
Quantitative processing Multipoint calibration curve with peak height/area/ratio, Multilinear regression (MLR method)
Spectrum search Search parameters, Search, Library creation
Print processing Report Generator
Optional Software Macro Programming, PLS Quantitation, Curve Fitting, 3 Dimensional Presentation with Mapping
Audit Trail Container function with stored interferogram/background spectrum, operation history
Password Protection
Log recording
FDA CFR Part 11 compliance, Electric Signature
Accessory Detection Automatically recognize Installed accessory. Then sets up scan parameters and runs macro program.
ATR Accessories; ATR-8000A, ATR-8200HA, MIRacle A, DuraSamplIR A, etc.
Diffuse Reflectance Accessories; DRS-8000A, etc.
Reflectance Accessories; SRM-8000A, RAS-8000A, etc.
Dimensions 600 (W) x 680 (L) x 290 (H) mm
Weight 54 Kg





© Shimadzu Europa GmbH | Albert-Hahn-Str. 6-10 | 47269 Duisburg | Tel: 0203-76870 | Fax: 0203-766625