Shim-pack GIST Series
Shim-pack GIST Series
Ultra-high inertness and high durability
Shim-pack GIST series columns are packed with newly developed high-purity porous spherical silica. Superior inertness not only improves analytical precision, but also increases column durability. Shim-pack GIST series is capable of tolerating a wider pH (1-10) situation, making it suitable for analysis of many different compounds.
C18, C8, Phenyl and NH2 columns are offered to cover an extended range of selectivity and meet diverse research and development needs.
Furthermore, Shim-pack GIST series is designed with various particle size ranges, which make it easy to transfer your analysis from ultra-high-speed to conventional methods.
Analysis Examples
Nine Pesticides on a Golf CourseConditions
| Column | Shim-pack GIST-HP C18 (150 mmL. x 3.0 mmI.D., 3 μm) |
| Eluent | A) 50 mM KH2PO4 (pH 3.5, H3PO4) B) CH3CN A/B = 60/40 - 4 min - 40/60 |
| Flow Rate | 1.5 mL/min |
| Col. Temp. | 40 °C |
| Detection | UV 235 nm |
| Injection Vol. | 5.0 μL |
SugarsConditions
| Column | Shim-pack GIST-HP C18 (150 mmL. x 3.0 mmI.D., 3 μm) |
| Eluent | A) 50 mM KH2PO4 (pH 3.5, H3PO4) B) CH3CN A/B = 60/40 - 4 min - 40/60 |
| Flow Rate | 1.5 mL/min |
| Col. Temp. | 40 °C |
| Detection | UV 235 nm |
| Injection Vol. | 5.0 μL |
Specifications
| Shim-pack GIST C18 |
Shim-pack GIST C8 |
Shim-pack GIST Phenyl |
Shim-pack GIST NH2 |
|
|---|---|---|---|---|
| Bonded Phase | Octadecyl Groups | Octyl Group | Phenyl Group | Aminopropyl Group |
| Analysis Mode | Reversed Phase | Reversed Phase | Reversed Phase | HILIC mode, Normal Phase |
| Features | Ultra-high inertness and high durability | Ultra-high inertness and high durability | Extremly strong π-π interactions | Sugar analysis |
| Particle Size (μm) | 2, 3, 5 | 2, 3, 5 | 2, 3, 5 | 3, 5 |
| Pore Size (nm) | 10 | 10 | 10 | 10 |
| Surface Area (m2/g) | 350 | 350 | 350 | 350 |
| Carbon Loading (%) | 14 | 8 | 10 | 7 |
| End Cap | ○ | ○ | - | - |
| Inertness | ▶▶▶▶▶ | ▶▶▶▶▶ | ▶▶▶▶▷ | - |
| pH Range | 1-10 | 1-10 | 2-7.5 | 2-7.5 |