Applications
High Accuracy Measurement with Minimized Detector Switchover Noise and Bump
Noise and bump caused by detector switchover are minimized to assure accurate measurement.

The transmission spectra of color filter and polyester film are shown in the left and right figures, respectively.
Noise or bump caused by the detector switchover range at 870 nm and 1650 nm is hardly observed.
Integrating Sphere for Deep Ultraviolet Measurement

Spectra with low noise can be obtained even for wavelengths near 190 nm, which are difficult to measure accurately using a regular spectrophotometer. The ability to measure spectra in this wavelength range is especially helpful when measuring semiconductor materials used for ArF excimer lasers.
The 100% baseline spectra measured on the SolidSpec-3700i DUV with an integrating sphere for the deep ultraviolet measurement and the SolidSpec-3700i with a normal integrating sphere are shown in the left figure.

In order to perform high-accuracy measurements in the deep UV region, a sufficient quantity of light and significant low stray light are required. The transmission spectrum of a silica plate measured with the Direct Detection Unit DDU-DUV(option) is shown in the left figure. Spectra with significantly lower noise are obtainable in the ultraviolet region.