Specifications
Specifications
Hardware Specifications
General Specifications
| Measurement principle | Laser Diffraction Method |
|---|---|
| Measurement range |
SALD-MS23: 17nm(0.017μm) to 2500μm |
| SALD-BC23: 17nm(0.017μm) to 400μm | |
| SALD-HC23: 30nm(0.030μm) to 280μm | |
| SALD-DS5: 300nm(0.3μm) to 2500μm |
Note 1: The measurement range varies according to the shape etc. of the particle.
Measurement Unit: SALD-2300
| Light source | Red Semiconductor laser (Wavelength 680 nm) |
|---|---|
| Light detector | Detector elements for UV semiconductor laser Total 84 elements (78 forward, 1 side, 5 back) |
| System Compliance | Class 1 Laser Product, CE compliant |
| Required power supply | 115 or 230 VAC as ordered 100 VA |
| Dimensions & weight | W680mm×D280mm×H430mm, 31kg |
| Operating Environment | Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Note 2: Reference sample and USB cable (2 m) supplied as standard
Sampler: SALD-MS23
| Dispersing bath | Capacity: 100~280cm3 |
|---|---|
| Sonicator | Frequency about 32 kHz, output about 40 W |
| Liquid Pump | Radial pump, maximum flow rate 2000cm3/min |
| Liquid Pump Material | Stainless (SUS 304, SUS 316), Tetrafluoroethylene (PTFE), Perfluoroelastmor (FEP) or Kalrez, Thermoflon Pascal (inside) |
| Liquid Supply Pump | Diaphragm pump, maximum flow rate 750cm3/min |
| Liquid Supply Pump Material | Tetrafluoroethylene, polyvinyl dene fluoride |
| Flow Cell | Quartz glass |
| Required power supply | 115 or 230 VAC as ordered, 200 VA |
| Dimensions & weight | W390mm×D520mm×H430mm,18kg |
| Operating Environment | Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Note 3: USB cable (2 m) supplied as standard
Batch Cell: SALD-BC23
| Cell Material | Quartz glass |
|---|---|
| Required Liquid Volume | Approx. 12 cm3 |
| Stirrer Mechanism | Up-and-down movement of blade |
| Dimensions & weight | W100mm×D120mm×H140mm, 0.8kg |
| Operating Environment | Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
High-concentration Sample Measurement System: SALD-HC23
| Cell Material | Borosilicate glass |
|---|---|
| Required Liquid Volume | Approx. 0.15 cm3 |
| Dimensions & weight | W20mm×D100mm×H9mm, 0.2kg |
| Operating Environment | Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Cyclone Injection Type Dry Measurement Unit: SALD-DS5
| Sample suction types | Cyclone type / One shot type / Hand shot type |
|---|---|
| Suction nozzle | Can be selected from 3 types |
| Sampling Unit Specifications | |
| System | Cyclone type |
| Communication method | USB (PC control) |
| Required power supply | 115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor) |
| Dimensions & weight | W 240×D 310×H 210 mm, 10 kg |
| Operating Environment | Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
| Pressure Regulator Specifications | |
| Primary pressure | 0.6 to 0.8 MPa |
| Secondary pressure | 0.05 to 0.5 MPa |
| Filtration rating | Removing particles of 5 μm or larger |
| Connection to air source | Tube of 6 mm outside diameter |
| Communication method | USB (PC control) |
| Required power supply | 115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor) |
| Dimensions & weight | W 130×D 223×H 233 mm, 3 kg |
| Operating Environment | Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation) |
Requirements for compressor and dust collector
| Compressor | Output | 0.4 kW | |
|---|---|---|---|
| Minimum pressure | Approx. 7 kgf/cm2(approx. 0.69 MPa) | ||
| Air discharge | 45 L/min | ||
| Tank volume | Approx. 30 L | ||
| Dust collector |
Type | Vacuum cleaner (Paper pack type) | |
| Dust collecting efficiency | More than 99% for 0.3 μm particles | ||
| Capacity | 2.0 m3/min or more | ||
| Vacuum | Approx. 2000 mm Aq or lower | ||
| Suction hose diameter | Approx. 32 mm | ||
Software Specifications
WingSALD II
| Output Conditions | |
|---|---|
| Particle Size (μm) Divisions | Fixed 51 or 101 divisions : User settable 51 divisions |
| Particle Amount (%) Divisions | Fixed 51 divisions : User settable 51 divisions |
| Distribution Basis | Count, length, area, or volume |
| Expression of Cumulative Distribution | Oversized or undersized |
| Expression of Frequency Distribution | q, q / Δ×, q / Δlog × |
| Smoothing Levels | 10 levels |
| Distribution Function Fitting | Rosin-Rammler distribution, logarithmic Gaussian distribution |
| Data Shifting | ±10 levels |
| Report Function | Single data sets (6 templates), overlaid data (5 templates), statistical data, time-series data, or 3D data can be selected and output using batch processing |
| Data Analysis Functions | |
| Scattering Angle Evaluation Function | Evaluates scattering characteristics within micro angle regions for samples such as optical films and sheets. |
| Data Emulation Functions | Emulates measurement results from other instruments and measurement principles, using SALD series measurement results. |
| Mixture Data Simulation Function | Simulates particle size distributions using any mixture ratio of multiple particle size distributions. |
| Data Connection Function | Combines two particle size distributions with different measurement ranges at any particle size point to create a single particle size distribution. |
| Continuous Measurement Function | Continuously measures changes in particle size distributions and particle diameters over time, at intervals as short as one second, and saves the results. |
| *1: | Windows is a registered trademark of Microsoft Corporation in the United States and other countries. |
| *2: | Other company names and product names mentioned in this document are registered trademarks or trademarks of their respective companies. |
| *3: | The TM and ® symbols are omitted in this document. |
* Appearance and specifications are subject to change without notice.