    LC-MS/MS Method Package for Short Chain Fatty Acids

    For LabSolutions™ LCMS Software

    Short-Chain Fatty Acids and Organic Acids Targeted for Analysis

    Of the short-chain fatty acids produced by intestinal bacteria, acetic acid, propionic acid, and butyric acid are well known, and it has been reported that there are some connections between them and lifestyle-related diseases such as obesity and diabetes. Generally speaking, short-chain fatty acids are highly volatile and highly hydrophilic. This makes it difficult to perform LCMS analysis using a conventional reverse phase system. For that reason, this method package targets short-chain fatty acids (C2 to C5) that have been derivatized using 3-nitrophenylhydrazine (3-NPH). After setting MRM transitions, it can be used for the simultaneous analysis (of 22 components), including organic acids related to the central metabolic pathways.

    Short-Chain Fatty Acids
    Acetic acid
    Propionic acid
    Butyric acid
    Isobutyric acid
    Valeric acid
    Isovaleric acid
    Organic Acids
    2-oxobutyric acid
    2-hydroxyglutaric acid
    α-ketoglutaric acid
    β-hydroxybutyric acid
    Isocitric acid
    Oxaloacetic acid
    Citric acid
    Glyoxylic acid
    Glycolic acid
    Succinic acid
    Lactic acid
    Pyruvic acid
    Fumaric acid
    Maleic acid
    Malonic acid
    Malic acid

    MRM Transition Settings

    When making MRM transition settings, selectivity can be enhanced by specifying characteristic product ions for 3-nitrophenylhydrazine derivatives. Note that some organic acids have a carbonyl group derived from ketones, so 3-nitrophenylhydrazine derivatives that reacted with carboxylic acid and the carbonyl group are targeted for MRM transition.

    Pretreatment Protocol Including Derivatization

    The entire protocol necessary for pretreatment, including the derivatization step using 3-nitrophenylhydrazine, is contained in the Instruction Manual for this method package. By following these procedures, an analyst should be able to perform the entire process, from derivatization to data acquisition and data analysis, immediately after the package is introduced.

    For the derivatization using 3-nitrophenylhydrazine, carbodiimide is used as a condensing agent, and reaction proceeds under mild conditions. Moreover, in order to efficiently carry out reactions with both short-chain fatty acids and organic acids, this method package includes derivatization using 3-NPH instead of the more commonly used 2-NPH. Also, 3-NPH derivatizations after reactions are stable, and quantitative analysis of the short-chain fatty acids or organic acids in the sample can be performed through reaction with an appropriate amount of derivatization reagent. Additionally, as reactions take place in water solutions using this derivatization reagent, the sample will not be dried and volatile components will not be lost.

    Derivatization Reaction with 3-nitrophenylhydrazine (3-NPH)

     
     
     

    The following example shows an analysis using the LCMS-8060 of a standard mixed solution of short-chain fatty acids/organic acids. An MRM chromatogram of short-chain fatty acids/organic acids derivatized using 3-NPH is shown. As can be seen, sufficient separations of the components, such as the structural isomers butyric acid, isobutyric acid, fumaric acid, maleic acid, citric acid, and isocitric acid were obtained. This method package includes MRM transitions that were formulated beforehand in order to achieve optimization for 3-NPH derivatizations for short-chain fatty acids/organic acids. As a result, short-chain fatty acids (C2 to C5) and organic acids representative of central metabolic pathways, which are difficult to analyze with LCMS using a conventional reversed phase system, can be analyzed simultaneously.

    chromatogram of short-chain fatty acids
     

    Remarks and Precautions

    LabSolutions LCMS Ver. 5.99SP2 or later is required.

     