Servopulser series systems allow selection of the optimal combination of units based on testing objectives. In addition, an extensive selection of optional testing equipment, such as various testing jigs, detectors, and atmospheric control testing units, is available. For more details, refer to the separate optional accessories brochure.

Tensile and Compression Test Jigs

Front-Opening Hydraulic Grip

Designed for full-amplitude tensile and compression fatigue testing, these grips offer superior ease-of-operation and ensure high-accuracy testing for a wide range of tests.

Maximum test force ±20 to 200 kN
(multiple capacities available)
Operating temperature range RT to +50 °C
Applicable sample Rod / flat plate

Manual Non-Shift Plate Grip

These grips are designed for full-amplitude tensile and compression fatigue testing of flat plate materials and feature a simple and efficient construction.

Maximum test force ±5 to 100 kN
(multiple capacities available)
Operating temperature range RT to +50 °C
-196 to +300 °C
Applicable sample Flat plate
 
 
 

Split Flange Rod Grip

These grips allow samples to be secured easily and firmly. They are ideal for full-amplitude tensile and compression fatigue testing of round rod samples.

Maximum test force ±10 to 200 kN
(multiple capacities available)
Operating temperature range RT to +100 °C
-196 to 300 °C
Applicable sample Rod
 
 
 

Pin-Type Grip for Flat Samples + Dynamic Strain Gauge
(for gauge length displacement)

These grips are designed for half-amplitude tensile fatigue testing.
Note: Supports only tensile testing.

Maximum test force +6 kN/10 kN
Operating temperature range -196 to +300 °C (±6 kN)
-20 to +300 °C (±10 kN)
Applicable sample Flat plate (max. 30 mm wide and 5 mm thick)
 
 
 

Non-Shift Wedge Grip for Static Testing

These grips can only be used for static testing. These high-capacity grips apply the self-tightening action of a wedge.
Note: Supports only tensile testing.

Maximum test force ±20 to 250 kN
(multiple capacities available)
Operating temperature range 0 to +120 °C
Applicable sample Rod / flat plate
 
 
 

Grips for CT Test Samples + Clip Gauge

These compact grips are designed specifically for tensile test samples and are compliant with ASTM E399 and E1820 standards. They can be used for tests performed to determine fracture toughness or crack propagation.
Note: Supports only tensile testing.

Maximum test force ±6 to 80 kN
Operating temperature range RT to +100 °C
-20 to 300 °C
Applicable sample CT test sample
 
 
 

Bolt Testing Jigs

Screw Tensile Test Jig

This jig is for tensile fatigue testing of various nuts and bolts. Various grips sizes are available depending on the bolt size.

Maximum test force ±100/250 kN
Operating temperature range RT to +50 °C
Applicable sample Nuts/bolts
 
 
 

Screw Looseness Test Device

This device allows testing various parameters to determine the loosening process of bolts. It applies a vibrational displacement in the thread tightening direction and in the perpendicular direction and then measures the change in tightening force in relation to the number of vibrations.

Maximum test force ±20 kN
Operating temperature range RT to +100 °C
-20 to +300 °C
Applicable sample CT test sample
 
 
 

Compression and Bending Test Jigs

Compression Plate

Compression plates are available with both he top and bottom fixed or with the top compression plate mounted on a spherical seat.

Maximum test force 20 to 500 kN
(multiple capacities available)
Operating temperature range RT to +250 °C
Applicable sample 60 to 220 mm dia.
 
 
 

3-Point/4-Point Bending Test Jig
(for partial half-amplitude fatigue testing)

Maximum test force 2kN
Max. dynamic bending moment 50 N/m
Operating temperature range RT to +100 °C
-196 to +300 °C
Jig dimensions Lower span: 30 to 100 mm
Upper span: 15 to 50 mm
 
 
 

Uniform Bending Test Jig (for full-amplitude fatigue testing)

This jig uses ball bearings at each support point to apply uniform bending loads.

Maximum test force ±2 to 10 kN
Max. dynamic bending moment ±20 to 250 N/m
Operating temperature range RT to +50 °C
-196 to +200 °C

CTOD Bending Test Jig + Clip Gauge

This jig is for CTOD bending tests compliant with ASTM E339. It is used for fracture toughness testing.

Maximum test force 50/100 kN
Max. dynamic bending moment 2/6 k N/m
Operating temperature range RT to +100 °C
-196 to +300 °C
 
 
 

Various Environmental Control Testing Systems

Various environmental control testing systems are required to simulate harsh environments or environments where materials are actually used, such as thermostatic, high-temperature, or extremely low-temperature environments. Therefore, an environmental control system can be added in the large testing space provided by the Servopulser series system.
See page 58.

{"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3578,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]} {"title":"Applications","source":"product","key":3578,"max":3,"filter_types":["applications","application_note","posters"],"link_title":"Learn more","link_url":false,"config_list":[],"page_links":[],"column_title":"Documents"} {"title":"Technical Documents","source":"product","key":3578,"max":3,"filter_types":["technical","technical_reports","white_papers","primers"],"link_title":"Learn more","link_url":"#table_anchor_technical","config_list":[],"page_links":[],"column_title":"Documents"} {"title":"Manuals","source":"product","key":3578,"max":3,"filter_types":["manuals"],"link_title":"Learn more","link_url":"#table_anchor_manual","config_list":[],"page_links":[],"column_title":"Documents"}