0

    eTOC series - Specs

    TOC-1000e On-Line TOC Analyzer for Pure Water

    Installation Space

    TOC-1000e

    TOC-1000e

    TOC-1000e with Vial Sampler (Optional)

    TOC-1000e with Vial Sampler (Optional)

    eTOC Series

    Specifications*

    Model

    TOC-1000e

    Measurement items

    TOC, conductivity (or specific resistance), and temperature

    Measurement principle

    UV oxidation-conductivity method

    Sample Water Conditions

    Temp.: 10 to 50 °C
    Flowrate: 30 to 500 mL/min
    Conductivity: 2 µS/cm or lower

    Measurement cycle

    2.5 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2 hrs, 4 hrs, 12 hrs, or 24 hrs

    Measurement range

    TOC: 0 to 2000 µg/L
    Conductivity: 0.023 to 206 µS/cm (without temperature correction)
    Temp.: 10 to 50 °C

    Detection limit

    TOC: 0.1 µg/L

    Accuracy

    TOC: ± 5 % (500 µg/L sucrose)
    Conductivity: ± 2 % (147 µS/cm at 25 °C)
    Temp.: ± 0.5 °C (at 25 °C)

    Repeatability

    TOC: CV ≤ 1 % or SD ≤ 0.3 µg/L, whichever is greater

    Linearity

    TOC: R2 ≥ 0.98

    Calibration*2

    Creates a 4-point calibration curve automatically

    Outputs

    USB flash drive, printer, LAN (file or Internet), Modbus (TCP/IP), analog*3, or contact*3

    Analog Output*3

    4 to 20 mA output, three channels from TOC, conductivity, specific resistance, or temperature

    Contact Output*3

    Alarm: Measurement value alarm, device alarm, serious failure, or power supply interruption
    Event: Ready for measurement, measuring, and analog synchronization signal

    Contact Input*3

    Start measurement

    Display

    7-inch color touch panel

    Ingress Protection Rating

    IP33

    Installation Site Requirements

    Indoors, with 10 to 40 °C ambient temperatures
    Tabletop, wall-mounted*4 or pole-mounted*4

    Power supply

    100 to 240 V AC, 50/60 Hz

    Dimensions

    W270 × D140 × H180 mm

    Weight

    Approx. 2.88 kg

    *1: The specification above is available when managed and operated appropriately by trained people.
    *2: Using an optional vial sampler
    *3: Using an optional external I/O kit
    *4: Using an optional bracket kit

    Read more