eTOC series - Specs
TOC-1000e On-Line TOC Analyzer for Pure Water
Installation Space
TOC-1000e

TOC-1000e with Vial Sampler (Optional)

eTOC Series
Specifications*
| Model |
TOC-1000e |
|---|---|
| Measurement items |
TOC, conductivity (or specific resistance), and temperature |
| Measurement principle |
UV oxidation-conductivity method |
| Sample Water Conditions |
Temp.: 10 to 50 °C |
| Measurement cycle |
2.5 min, 5 min, 10 min, 15 min, 30 min, 1 hr, 2 hrs, 4 hrs, 12 hrs, or 24 hrs |
| Measurement range |
TOC: 0 to 2000 µg/L |
| Detection limit |
TOC: 0.1 µg/L |
| Accuracy |
TOC: ± 5 % (500 µg/L sucrose) |
| Repeatability |
TOC: CV ≤ 1 % or SD ≤ 0.3 µg/L, whichever is greater |
| Linearity |
TOC: R2 ≥ 0.98 |
| Calibration*2 |
Creates a 4-point calibration curve automatically |
| Outputs |
USB flash drive, printer, LAN (file or Internet), Modbus (TCP/IP), analog*3, or contact*3 |
| Analog Output*3 |
4 to 20 mA output, three channels from TOC, conductivity, specific resistance, or temperature |
| Contact Output*3 |
Alarm: Measurement value alarm, device alarm, serious failure, or power supply interruption |
| Contact Input*3 |
Start measurement |
| Display |
7-inch color touch panel |
| Ingress Protection Rating |
IP33 |
| Installation Site Requirements |
Indoors, with 10 to 40 °C ambient temperatures |
| Power supply |
100 to 240 V AC, 50/60 Hz |
| Dimensions |
W270 × D140 × H180 mm |
| Weight |
Approx. 2.88 kg |
*1: The specification above is available when managed and operated appropriately by trained people.
*2: Using an optional vial sampler
*3: Using an optional external I/O kit
*4: Using an optional bracket kit