Autograph AGS-X HC Series - Specs

AUTOGRAPH Precision Universal Tester

AGS-50kNXD

Specifications Table-top

Max. Load Capacity

50kN

Lineup of Models Differing byLoad Cell Capacity

50kN

Loading Method

Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive

Force : Accuracy : High-Precision Type
(1/500, ±0.5%)

Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5,
BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5. *1

Measurement : Accuracy : Standard-Precision Type (1/500, ±1%)

Within ±1% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 1, ISO 7500-1 Class 1,
BS 1610 Class 1, and ASTM E4, JIS B7721 Class 1. *1

Measurement : Calibration

Automatic test force calibration: select tensile, compression, or
tensile and compression

Crosshead : Speed Range

0.001 to 800 mm/min (stepless)

Crosshead : Max. Return Speed

1100mm/min

Crosshead Speed
Accuracy*2

±0.1

Crosshead Speed and
Permitted Test Force

To load cell capacity across entire speed range

Crosshead – Table
Distance(Tensile stroke) *3

1210mm
(745mm, MWG)

Effective Test Width

425mm

Crosshead
Position
Detection : Measurement Method

Optical encoder

Crosshead
Position
Detection : Display Method

Digital display (display resolution: 0.001 mm)

Crosshead
Position
Detection : Positional Accuracy

±0.1% indicated value or ±0.01 mm, whichever is larger

Sampling Speed

1000Hz max. *4

Standard Functions

• Automatic reading of load cell characteristic values
• Test force display, stress display, stroke display,position display
• External analog output (2 channels)

Standard Functions

• External analog input (2 channels) *4
• External digital input (2 channels) *4

Standard Functions

• Analog recorder (option) output
• Dataletty (option) output*5
• Automatic test force / stress control (Autotuning)
• Automatic strain control (Autotuning) *4
• Test force auto-zeroing
• Test force auto-calibration
• Break detection, auto-return
• Load cell overload detection
• Touch-load detection function

Accessories

Load Cell (with CAL cable), Power cable (5 m), turning rod, cable clamps, instruction manual

Dimensions

W718×D641×
H1633mm

Weight

260kg

Power Requirements

Single phase AC200-230V 50/60Hz 4.0kVA

Power Requirements

Supply voltage fluctuations within ±10% of the set value.
D-class (100 Ω max.) grounding resistance.

Operating Environment

Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation)
Floor vibrations: frequency 10 Hz max., amplitude 5 μm max.

*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation

Read more
Show Less

AGS-100kNX

Specifications Floor Type

Capacity

100kN

Loading Method

Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive

Force
Measurement
: Accuracy : High-Precision Type(1/500, ±0.5%)

Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5, BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5. *1

Force
Measurement
: Accuracy : Standard-Precision Type(1/500, ±1%)

Within ±1% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 1, ISO 7500-1 Class 1, BS 1610 Class 1, and ASTM E4, JIS B7721 Class 1. *1

Force
Measurement
: Calibration

Automatic test force calibration: select tensile, compression, or tensile and compression

Crosshead : Speed Range

0.001 to 800 mm/min (stepless)

Crosshead : Max. Return Speed

1100mm/min

Crosshead Speed Accuracy *2

±0.1%

Crosshead Speed and Permitted Test Force

To load cell capacity across entire speed range

Crosshead - Table Distance
(Tensile stroke) *3

1255mm (745mm, MWG)

Effective Test Width

600mm

Crosshead
Position
Detection : Measurement Method

Optical encoder

Crosshead
Position
Detection : Display Method

Digital display (display resolution: 0.001 mm)

Crosshead
Position
Detection : Positional Accuracy

±0.1% indicated value or ±0.01 mm, whichever is larger

Sampling Speed

1000Hz max. *4

Test Method Files

40 files (PC link: 20 files, standalone controller: 20 files)

Standard Functions

• Automatic reading of load cell characteristic values
• Test force display, stress display, stroke display,position display
• External analog output (2 channels)

Standard Functions

• External analog input (2 channels) *4

Standard Functions

• External digital input (2 channels) *4
• Analog recorder (option) output
• Dataletty (option) output *5
• Automatic test force / stress control (Autotuning)
• Automatic strain control (Autotuning) *4
• Test force auto-zeroing
• Test force auto-calibration
• Break detection, auto-return
• Load cell overload detection
• Touch-load detection function

Accessories

Load cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual

Dimensions

W945×D725×H2164mm

Weight

525kg

*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation

Read more
Show Less

AGS-300kNX

Specifications Floor Type

Capacity

300kN

Loading Method

Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive

Force
Measurement
: Accuracy : High-Precision Type(1/500, ±0.5%)

Within ±0.5% indicated test force (at 1/250 to 1/1 load cell rating) 
Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5, BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5. *1

Force
Measurement
: Accuracy : Standard-Precision Type(1/500, ±1%)

Within ±1% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 1, ISO 7500-1 Class 1, BS 1610 Class 1, and ASTM E4, JIS B7721 Class 1.*1

Force
Measurement
: Calibration

Automatic test force calibration: select tensile, compression, or tensile and compression

Crosshead : Speed Range

0.001 to 500 mm/min (stepless)

Crosshead : Max. Return Speed

550mm/min

Crosshead Speed Accuracy *2

±0.1%

Crosshead Speed and Permitted Test Force

0.001 to 200kN:0.001 to 500mm/min(entire speed range) 200kN to 300kN:0.001 to 400mm/min

Crosshead - Table Distance
(Tensile stroke) *3

1475mm (635mm, MWG)

Effective Test Width

600mm

Crosshead
Position
Detection : Measurement Method

Optical encoder

Crosshead
Position
Detection : Display Method

Digital display (display resolution: 0.001 mm)

Crosshead
Position
Detection : Positional Accuracy

±0.1% indicated value or ±0.01 mm, whichever is larger

Sampling Speed

1000Hz max. *4

Test Method Files

40 files (PC link: 20 files, standalone controller: 20 files)

Standard Functions

• Automatic reading of load cell characteristic values
• Test force display, stress display, stroke display,position display
• External analog output (2 channels)

Standard Functions

• External analog input (2 channels) *4

Standard Functions

• External digital input (2 channels) *4
• Analog recorder (option) output
• Dataletty (option) output *5
• Automatic test force / stress control (Autotuning)
• Automatic strain control (Autotuning) *4
• Test force auto-zeroing
• Test force auto-calibration
• Break detection, auto-return
• Load cell overload detection
• Touch-load detection function

Accessories

Load cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual

Dimensions

W945×D725×H2414mm

Weight

675kg

*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation

Read more
Show Less

AGS-20kNXD

Specifications Table-top

Max. Load Capacity

20kN

Lineup of Models Differing byLoad Cell Capacity

20kN

Loading Method

Direct high-precision constant speed strain control method via non-backlash ball-screw drive

Force:Accuracy : High-Precision Type
(1/500, ±0.5%)

Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 0.5, ISO 7500-1 Class 0.5,
BS 1610 Class 0.5, and ASTM E4, JIS B7721 Class 0.5. *1

Measurement : Accuracy:Standard-Precision Type (1/500, ±1%)

Within ±1% indicated test force (at 1/500 to 1/1 load cell rating)
Conforms to EN 10002-2 Grade 1, ISO 7500-1 Class 1,
BS 1610 Class 1, and ASTM E4, JIS B7721 Class 1. *1

Measurement : Calibration

Automatic test force calibration: select tensile, compression, or
tensile and compression

Crosshead : Speed Range

0.001 to 1600 mm/min (stepless)

Crosshead : Max. Return Speed

2200mm/min

Crosshead Speed
Accuracy*2

±0.1

Crosshead Speed and
Permitted Test Force

To load cell capacity across entire speed range

Crosshead – Table
Distance(Tensile stroke) *3

1250mm
(765mm, MWG)

Effective Test Width

425mm

Crosshead
Position
Detection:Measurement Method

Optical encoder

Crosshead
Position
Detection:Display Method

Digital display (display resolution: 0.001 mm)

Crosshead
Position
Detection:Positional Accuracy

±0.1% indicated value or ±0.01 mm, whichever is larger

Sampling Speed

1000Hz max. *4

Standard Functions

• Automatic reading of load cell characteristic values
• Test force display, stress display, stroke display,position display
• External analog output (2 channels)

Standard Functions

• External analog input (2 channels) *4
• External digital input (2 channels) *4

Standard Functions

• Analog recorder (option) output
• Dataletty (option) output*5
• Automatic test force / stress control (Autotuning)
• Automatic strain control (Autotuning) *4
• Test force auto-zeroing
• Test force auto-calibration
• Break detection, auto-return
• Load cell overload detection
• Touch-load detection function

Accessories

Load Cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual

Dimensions

W718×D641×
H1633mm

Weight

235kg

Power Requirements

Single phase AC200-230V 50/60Hz 4.0kVA

Power Requirements

Supply voltage fluctuations within ±10% of the set value.
D-class (100 Ω max.) grounding resistance.

Operating Environment

Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation)
Floor vibrations: frequency 10 Hz max., amplitude 5 μm max.

*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
* Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
* Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation

Read more
Show Less
{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":3432,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":3432,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}