Specifications of AG-Xplus floor type

LCD touch panel, laptop PC, and table are optional.
Installation Space
(Dimensions given for left, right, and back of main unit are required for maintenance.)

AG-Xplus Model Capacities and Kit Numbers
Kit number (AG-Xplus unit + load cell set + upper/lower joint)

| Model | Weight (approx. kg) |
Power Requirements Value in ( ) indicates power consumption | Installation Environment |
|---|---|---|---|
| AG-20kN/50kNXplus | 654 | 3-phase, 200 to 230 V, 50 to 60 Hz, 5 kVA (1.2 kW) |
Temp.: 5 to 40 °C Humidity: 20 to 80 %, (no condensation) Voltage fluctuation: ±10 % max. Vibration: Frequency 10 Hz max., amplitude 5 mm max. |
| AG-100kNXplus | 834 | 3-phase, 200 to 230 V, 50 to 60 Hz, 7 kVA (2.0 kW) | |
| AG-250kN/300kNXplus | 960 | 3-phase, 200 to 230 V, 50 to 60 Hz, 7.5 kVA (2.5 kW) |
Note) Requires Class D grounding (with 100 ohm max. ground resistance).

Floor Type AG-Xplus
*1: Crosshead speed precision is calculated using crosshead transfer amount within a specified period of time for
the crosshead speed of 0.5 mm/min to 500 mm/min under normal conditions.
*2: Tensile stroke is the value used when attaching the MWG(non-shift wedge type)grips.
Stroke can be extended.
Values under 5kN are with SCG(screw type flat)grips attached.
*3: JIS B7721, EN 10002-2, ISO 7500-1, and ASTM E4 standards recommend re-verification after installation of
testing machine.
*4: The LCD touch panel(optional)is required for Dataletty(optional).
*5: Dateletty(optional)and TRAPEZIUM X operational software cannot be used together.
Values in this table have been measured based on separately-approved test standards.
Extensions to the Main Unit (Floor Type and Table-Top Type)
Wide Frame Series
975, 1100, and 1375 mm effective width models are available. Use these models for testing large, actual object specimens.
Reinforced Yoke Series
Use for testing between the crosshead and yoke.
Ultralow-Speed Crosshead Series
Extends crosshead speeds into the ultralow-speed range.
Large Capacity Series
These are 2000 kN, 1000 kN, 600 kN, 500 kN large-capacity testing machines demanded by heavy industry, including steel, construction, and ship-building. These machines are custom-made.
Extended Column Series
Models are available with the clearance between the yoke and table extended to 250 mm, 500 mm, or 750 mm. Consider these machines for testing materials requiring long tensile strokes.
6000 mm/min High-Speed Return Series
Machines can be custom-made with 6000 mm/min or greater crosshead speed.
| Ultralow-Speed Series | |
|---|---|
|
Speed Range (mm/min) |
0.00005 to 1000 However, up to 500 for 250 kN and 300 kN models |