AUTOGRAPH AGS-V Series - Specs
Tabletop Precision Universal Testing Machines
Specifications
Model Name | AGS-1NVD/ 2NVD/5NVD |
AGS-10NVD/ 20NVD/50NVD |
AGS- 100NVD |
AGS- 500NVD |
AGS- 1kNVD |
AGS- 5kNVD |
AGS- 10kNVD |
|||||
Load Capacity | 1N/2N/5N | 10N/20N/50N | 100N | 500N | 1kN | 5kN | 10kN | |||||
Loading Method | Direct high-precision constant-speed strain control method using precision ball screw drive with no backlash | |||||||||||
Test Force |
Accuracy | High-Precision Type 1/500 | Within ±0.5 % of the displayed test force (within the range of 1/1 to 1/500 of the rated load cell capacity) JIS B7721 class 0.5, EN 10002-2 grade 0.5, ISO 7500-1 class 0.5, BS 1610 class 0.5, complies with ASTM E4 *1) |
|||||||||
Wide-Range Accuracy Type 1/1000 | Within ±1 % of the displayed test force (within the range of 1/1 to 1/1000 of the rated load cell capacity) JIS B7721 class 1, EN 10002-2 grade 1, ISO 7500-1 class 1, BS 1610 class 1, complies with ASTM E4*1) |
|||||||||||
Standard Accuracy Type 1/500 | Within ±1 % of the displayed test force (within the range of 1/1 to 1/500 of the rated load cell capacity) JIS B7721 class 1, EN 10002-2 grade 1, ISO 7500-1 class 1, BS 1610 class 1, complies with ASTM E4*1) |
|||||||||||
Test Force Calibration | Select from automatic test force calibration, tension, or tension and compression. | |||||||||||
Crosshead | Speed Range | 0.0005 to 1,500 mm/min (can be set as desired, not in steps) | ||||||||||
Maximum Return Speed | 1,650 mm/min | |||||||||||
Crosshead Speed Accuracy *2) | ±0.1% | |||||||||||
Crosshead Speed and Allowable Test Force | Up to the load cell capacity for all speeds | |||||||||||
Distance between Crosshead and Table (Tensile stroke) *3) | - | 1,180 mm (945 mm, SCG) |
1,180 mm (920 mm, SCG) |
1,180 mm (915 mm, SCG) |
1,180 mm (910 mm, SCG) |
1,180 mm (760 mm, MWG) |
1,180 mm (740 mm, MWG) |
|||||
Effective Test Width | 425mm | |||||||||||
Crosshead Position Detection | Measurement Method | Optical encoder | ||||||||||
Display Method | Digital display (display resolution: 0.001 mm) | |||||||||||
Positional Accuracy | Within ±0.1 % or within ±0.01 mm of indicated value, whichever is greater | |||||||||||
Sampling Rate | Maximum 0.2 msec | |||||||||||
Test Condition Files | 40 files | |||||||||||
Standard Functions |
|
|||||||||||
Accessories | Load Cell | 1 N/2 N/5 N | 10 N/20 N/50 N | 100 N | 500 N | 1 kN | 5 kN | 10 kN | ||||
Other | Power supply cable (2.4 m), rotation rod, DVD media (manual and other electronic data), Safety precautions sheet, Operation controller, Safety glasses, Safety gloves | |||||||||||
Size, Weight |
Standard Model | W657 mm×D525 mm×H1,562 mm Approx. 80 kg (100 V), 75 kg (200 V) |
||||||||||
250 mm Extension Model | W657 mm×D525 mm×H1,812 mm Approx. 85 kg (100 V), 80 kg (200 V) |
|||||||||||
500 mm Extension Model | W657 mm×D525 mm×H2,062 mm Approx. 90 kg (100 V), 85 kg (200 V) |
|||||||||||
Power Supply Requirements | 1φ 100 to 115 V (100 V), 1φ 200 to 230 V (200 V) 50/60 Hz 900 VA (100 V), 1.2 kVA (200 V) Power supply voltage fluctuations shall be within ±10 % of the installation value, and type D grounding (100 Ω or less) shall be provided. |
|||||||||||
Power Consumption | 100 V:650 W、200 V:600 W |
- In the JIS B7721, EN 10002-2, ISO 7500-1, ASTM E4 standards, testing is recommended after installation of the testing machine.
- The crosshead speed accuracy is calculated from the amount of crosshead movement within a prescribed time under steady state conditions of 0.5 mm/min to 500 mm/min.
- The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) grips are installed.
- There are options and jigs that cannot be used in conjunction with the protection cover. For more details, please contact us.