SALD-BC Series

Measurement is possible with small amounts of sample. Use of a funnel improves operation ease when pouring samples via the funnel, and the vertical motion of the stirring plate suppresses the adverse affects of sedimentation of particles to ensure accurate measurement. This is ideal for measurement of particles in the sub-micron region or lower. Also, waste solutions can be kept to a minimum when using organic solvents as the dispersion medium.
Features
Measurement Possible by Small Amounts of Samples (Measured Particles) and Medium Liquid (Dispersion Medium)
When performing circulation measurement, all of the flow cell, dispersing bath (sampler) and circulation tubing must be filled with suspension (sample and dispersion medium). When using a laser diffraction particle size analyzer, the sample concentration needed for measurement is about 100 ppm (wt%). Thus, the amount of sample may appear to be not so large, but some customers do have a need to perform measurement using even smaller amounts especially when measuring valuable and expensive samples. Also, some customers who use organic solvents as the dispersion medium want to use less suspension for measurement so that they can reduce waste solutions.
Particle Sedimentation Minimized by Vertical Motion of Stirring Plate

Fig. 1 Structure of Batch Cell
With a batch cell, sedimentation of particles would affect measurement since the suspension (sample and dispersion medium) is not circulated and is in a still state. Since larger sizes of particles sediment more quickly, the particle size distribution might become smaller over time. Since measurement durations are short on the SALD series, the affect of this is considered to be relatively small. Attention, however, is required when measuring samples having a wide distribution width.
The batch cell for the SALD series uses the vertical motion (see Fig. 1) of the stirring plate to suppress sedimentation of particles and ensure accurate measurement of particle size distribution. There is also no adverse affect arising from use of a stirrer when measuring magnetized particles since it does not incorporate a magnet unlike a magnetic stirrer.
Funnel Lessens the Possibility of Spilled Samples
A tetrafluoroethylene resin funnel (see Fig. 1) is provided to avoid suspension (sample and dispersion medium) being spilled outside when pouring it into the batch cell. This lessens the possibility of suspension getting on your hands and fingers. A major advantage of this is demonstrated when acid or organic solvents are used as the dispersion medium. When solvent transfers to the outer side of a batch cell, appropriate scattered light cannot be detected, which causes measurement error. Use of a funnel solves this problem.
Note
When particles of 100 µm or less are measured, the affect of sampling error increases if the distribution width is particularly wide since the number of particles in an appropriate concentration decreases.
The affect of sedimentation of particles also increases when particles of 100 µm or more are measured and the distribution width is wide. With large particle densities, the affect of sedimentation must also be taken into consideration at also particle sizes of approximately several µm.
In cases such as these, circulation measurement using a sampler is recommended.